DescriptionDräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.