Dräger Infinity M300 VG2.3.1 Network-Based Denial of Service

Go Back
severity
high
date
Affecting
  • Infinity M300 <= VG2.3.1 and earlier

CWE
  • CWE-400 Uncontrolled Resource Consumption
CVSS
7.1
CVSS V4 Vector
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
Description
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.