Advisories

Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling

Go Back
severity
high
date
Affecting
  • Infinity Acute Care System <= VG4.1.1, VG4.0.3, lower than VG4.0.3

  • Standalone Infinity M540 patient monitor <= VG4.1.1, lower than VG4.1.1

CWE
  • CWE-924 Improper Enforcement of Message Integrity During Transmission in a Communication Channel
CVSS
8.8
CVSS V4 Vector
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Ready to get Started?

Explore VulnCheck, a next-generation Cyber Threat Intelligence platform, which provides exploit and vulnerability intelligence to help you prioritize and remediate vulnerabilities that matter.
  • Vulnerability Prioritization
    Prioritize vulnerabilities that matter based on the threat landscape and defer vulnerabilities that don't.
  • Early Warning System
    Real-time alerting of changes in the vulnerability landscape so that you can take action before the attacks start.